Computed tomography (CT), originally known as computed axial tomography (CAT or CAT scan) and body section roentgenography, is a medical imaging method employing tomography where digital geometry processing is used to generate a three-dimensional image of the internals of an object from a large series of two-dimensional X-ray images taken around a single axis of rotation. The word "tomography" is derived from the Greek tomos (slice) and graphia (describing). CT produces a volume of data which can be manipulated, through a process known as windowing, in order to demonstrate various structures based on their ability to block the x-ray beam. Although historically (see below) the images generated were in the axial or transverse plane (orthogonal to the long axis of the body), modern scanners allow this volume of data to be reformatted in various planes or even as volumetric (3D) representations of structures.
Although most common in healthcare, CT is also used in other fields, e.g. nondestructive materials testing.
Allan McLeod Cormack of Tufts University independently invented a similar process and they shared a Nobel Prize in medicine in 1979. The original 1971 prototype took 160 parallel readings through 180 angles, each 1° apart, with each scan taking a little over five minutes. The images from these scans took 2.5 hours to be processed by algebraic reconstruction techniques on a large computer.
In the US, the machine sold for about $390,000, with the first installations being at the Lahey Clinic, then Massachusetts General Hospital, and George Washington University in 1973.
The first CT system that could make images of any part of the body, and did not require the "water tank" was the ACTA scanner designed by Robert S. Ledley, DDS at Georgetown University.
Bulky, expensive and fragile photomultiplier tubes gradually gave way to improved detectors. A xenon gas ionization chamber detector array was developed for third generation scanners, which provided greater resolution and sensitivity. Eventually, both of these technologies were replaced with solid-state detectors: rectangular, solid-state photodiodes, coated with a fluorescent rare earth phosphor. Solid state detectors were smaller, more sensitive and more stable, and were suitable for 3rd and 4th generation designs.
On an early 4th generation scanner, 600 photomultiplier tubes, 1/2 in (12 mm) in diameter, could fit in the detector ring. Three photodiode units could replace one photomultiplier tube. This change resulted in increasing both the acquisition speed, and image resolution. The method of scanning was still slow, because the X-ray tube and control components interfaced by cable, limiting the scan frame rotation.
Initially, 4th generation scanners carried a significant advantage - the detectors could be automatically calibrated on every scan. The fixed geometry of 3rd generation scanners was especially sensitive to detector mis-calibration (causing ring artifacts). Additionally, because the detectors were subject to movement and vibration, their calibration could drift significantly.
All modern medical scanners are of 3rd generation design. Modern solid-state detectors are sufficiently stable that calibration for each image is no longer required. The 4th generation scanners' inefficient use of detectors made them considerably more expensive than 3rd generation scanners. Further, they were more sensitive to artifacts because the non-fixed relationship to the x-ray source made it impossible to reject scattered radiation.
Slip-ring technology replaced the spooled cable technology of older CT scanners, allowing the X-ray tube and detectors to spin continuously. When combined with the ability to move the patient continuously through the scanner this refinement is called Helical CT or, more commonly, Spiral CT.
Multi-detector-row CT systems further accelerated scans, by allowing several images to be acquired simultaneously. Modern scanners are available with up to 64 detector rows / output channels ( depends upon the technology used by the manufacturer ). It is possible to complete a scan of the chest in a few seconds. An examination that required 10 separate breath-holds of 10 seconds each, can now be completed in a single 10 second breath-hold. Multi-detector CT can also provide isotropic resolution, permitting cross-sectional images to be reconstructed in arbitrary planes; an ability similar to MRI. More anatomical volume coverage in less time is one of the key feature of the latest generation MD CT Scanners. It is however more important to achieve better spatial resolution than only volume coverage for better reconstructed images. Latest generation MD CT scanners with flying X-Ray tube focal spot in z-axis direction shows better image resolution.
A different approach was used for a particular type of dedicated cardiac CT technique called electron-beam CT (also known as ultrafast CT, and occasionally fifth generation CT). With temporal resolution of approximately 50 ms, these scanners could freeze cardiac and pulmonary motion providing high quality images. Only one manufacturer offered these scanners (Imatron, later GE healthcare), and few of these scanners were ever installed, primarily due to the very high cost of the equipment and their single-purpose design. Rapid development of MDCT has significantly reduced the advantage of EBCT over conventional systems. Contemporary MDCT systems have temporal resolution approaching that of EBCT, but at lower cost and with much higher flexibility. Because of this, MDCT is usually the preferred choice for new installations.
Improved computer technology and reconstruction algorithms have permitted faster and more accurate reconstruction. On early scanners reconstruction could take several minutes per image, a modern scanner can reconstruct a 1000 image study in under 30 seconds. Refinements to the algorithms have reduced artifacts.
Dual source CT uses 2 x-ray sources and 2 detector arrays offset at 90 degrees. This reduces the time to acquire each image to about 0.1 seconds, making it possible to obtain high quality images of the heart without the need for heart rate lowering drugs such as beta blockers. A dual-source multi-detector row scanner can complete an entire cardiac study within a single 10 second breath hold.
Volumetric CT is an extension of multi-detector CT, currently at research stage. Current MDCT scanners sample a 4 cm wide volume in one rotation. Volumetric CT aims to increase the scan width to 10-20 cm, with current prototypes using 256 detector-rows. Potential applications include cardiac imaging (a complete 3D dataset could be acquired in the time between 2 successive beats) and 3D cine-angiography.
Newer machines with faster computer systems and newer software strategies can process not only individual cross sections but continuously changing cross sections as the gantry, with the object to be imaged, is slowly and smoothly slid through the X-ray circle. These are called helical or spiral CT machines. Their computer systems integrate the data of the moving individual slices to generate three dimensional volumetric information (3D-CT scan), in turn viewable from multiple different perspectives on attached CT workstation monitors.
In conventional CT machines, an X-Ray tube and detector are physically rotated behind a circular shroud (see the image above right); in the electron beam tomography (EBT) the tube is far larger and higher power to support the high temporal resolution. The electron beam is deflected in a hollow funnel shaped vacuum chamber. Xray is generated when the beam hits a stationary target. The detector is also stationary.
The data stream representing the varying radiographic intensity sensed reaching the detectors on the opposite side of the circle during each sweep— 360 or just over 180 degrees in conventional machines, 220 degree in EBT —is then computer processed to calculate cross-sectional estimations of the radiographic density, expressed in Hounsfield units.
CT is used in medicine as a diagnostic tool and as a guide for interventional procedures. Sometimes contrast materials such as intravenous iodinated contrast are used. This is useful to highlight structures such as blood vessels that otherwise would be difficult to delineate from their surroundings. Using contrast material can also help to obtain functional information about tissues.
Pixels in an image obtained by CT scanning are displayed in terms of relative radiodensity. The pixel itself is displayed according to the mean attenuation of the tissue(s) that it corresponds to on a scale from -1024 to +3071 on the Hounsfield scale. The phenomenon that one part of the detector can not differ between different tissues is called as Partial Volume Effect. That means that a big amount of cartilage and a thin layer of compact bone can cause the same attenuation in a voxel as hyperdens cartilage alone. Water has an attenuation of 0 Hounsfield units (HU) while air is -1000 HU, cancellous bone is typically +400 HU, cranial bone can reach 2000 HU or more (os temporale) and can cause artefacts. The attenuation of metallic implants depends on atomic number of used element: Titanium usually has an amount of +1000 HU, iron steel can completly extinguish the X-ray and is therefore responsible for well-known line-artefacts in computed tomogrammes.
Windowing is the process of using the calculated Hounsfield units to make an image. The various radiodensity amplitudes are mapped to 256 shades of gray. These shades of gray can be distributed over a wide range of HU values to get an overview of structures that attenuate the beam to widely varying degrees. Alternatively, these shades of gray can be distributed over a narrow range of HU values (called a narrow window) centered over the average HU value of a particular structure to be evaluated. In this way, subtle variations in the internal makeup of the structure can be discerned. This is a commonly used image processing technique knowna s contrast compression. For example, to evaluate the abdomen in order to find subtle masses in the liver, one might use liver windows. Choosing 70 HU as an average HU value for liver, the shades of gray can be distributed over a narrow window or range. One could use 170 HU as the narrow window, with 85 HU above the 70 HU average value; 85 HU below it. Therefore the liver window would extend from -15 HU to +155 HU. All the shades of gray for the image would be distributed in this range of Hounsfield values. Any HU value below -15 would be pure black, and any HU value above 155 HU would be pure white in this example. Using this same logic, bone windows would use a wide window (to evaluate everything from fat-containing medullary bone that contains the marrow, to the dense cortical bone), and the center or level would be a value in the hundreds of Hounsfield units.
First, CT completely eliminates the superimposition of images of structures outside the area of interest. Second, because of the inherent high-contrast resolution of CT, differences between tissues that differ in physical density by less than 1% can be distinguished. Third, data from a single CT imaging procedure consisting of either multiple contiguous or one helical scan can be viewed as images in the axial, coronal, or sagittal planes, depending on the diagnostic task. This is referred to as multiplanar reformatted imaging.
The greatly increased availability of CT, together with its value for an increasing number of conditions, has been responsible for a large rise in popularity. So large has been this rise that, in the most recent comprehensive survey in the UK, CT scans constituted 7% of all radiologic examinations, but contributed 47% of the total collective dose from medical X-ray examinations in 2000/2001 (Hart & Wall, European Journal of Radiology 2004;50:285-291). Increased CT usage has led to an overall rise in the total amount of medical radiation used, despite reductions in other areas.
The radiation dose for a particular study depends on multiple factors: volume scanned, patient build, number and type of scan sequences, and desired resolution and image quality.
| Examination | Typical effective dose (mSv) |
|---|---|
| Chest X-ray | 0.02 |
| Head CT | 1.5(a) |
| Abdomen | 5.3(a) |
| Chest | 5.8(a) |
| Chest, Abdomen and Pelvis | 9.9(a) |
| Cardiac CT angiogram | 6.7-13(b) |
| CT colongraphy (virtual colonoscopy) | 3.6 - 8.8 |
For detection of tumors, CT scanning with IV contrast is occasionally used but is less sensitive than magnetic resonance imaging (MRI).
CT can also be used to detect increases in intracranial pressure, e.g. before lumbar puncture or to evaluate the functioning of a ventriculoperitoneal shunt.
CT is also useful in the setting of trauma for evaluating facial and skull fractures.
In the head/neck/mouth area, CT scanning is used for surgical planning for craniofacial and dentofacial deformities, evaluation of cysts and some tumors of the jaws/sinuses/nasal cavity/orbits, diagnosis of the causes of chronic sinusitus, and for planning of dental implant reconstruction.
CT is excellent for detecting both acute and chronic changes in the lung parenchyma. For detection of airspace disease (such as pneumonia) or cancer, ordinary non-contrast scans are adequate.
For evaluation of chronic interstitial processes (emphysema, fibrosis, and so forth), thin sections with high spatial frequency reconstructions are used. For evaluation of the mediastinum and hilar regions for lymphadenopathy, IV contrast is administered.
CT angiography of the chest (CTPA) is also becoming the primary method for detecting pulmonary embolism (PE) and aortic dissection, and requires accurately timed rapid injections of contrast and high-speed helical scanners. CT is the standard method of evaluating abnormalities seen on chest X-ray and of following findings of uncertain acute significance.
Cardiac MSCT carries very real risks since it exposes the subject to the equivalent of 500 chest X Rays in terms of radiation. The relationship of radiation exposure to increased risk in breast cancer has yet to be definitively explored.
Also a lot of MSCT technicians are trained cardiologist as opposed to radiologists. The positive (93-95%) and negative (97-98%) predictive values of the scan are calculated on the basis of a knowledgable staff which may not always be the case.
Much of the software is based on data findings from caucasian study groups and as such the assumptions made may also not be totally true for all other populations.
Dual Source CT scanners, introduced in 2005, allow higher temporal resolution therefore reducing the scan time for one heart scan. Scan time can be an important factor, because the examination needs to be performed during breath hold. A critical ill patient might not be able to hold the breath long enough.
Oral and/or rectal contrast may be used depending on the indications for the scan. A dilute (2% w/v) suspension of barium sulfate is most commonly used. The concentrated barium sulfate preparations used for fluoroscopy e.g. barium enema are too dense and cause severe artifacts on CT. Iodinated contrast agents may be used if barium is contraindicated (e.g. suspicion of bowel injury). Other agents may be required to optimize the imaging of specific organs: e.g. rectally administered gas (air or carbon dioxide) for a colon study, or oral water for a stomach study.
CT has limited application in the evaluation of the pelvis. For the female pelvis in particular, ultrasound is the imaging modality of choice. Nevertheless, it may be part of abdominal scanning (e.g. for tumors), and has uses in assessing fractures.
CT is also used in osteoporosis studies and research along side DXA scanning. Both CT and DXA can be used to asses bone mineral density (BMD) which is used to indicate bone strength, however CT results do not correlate exactly with DXA (the gold standard of BMD measurment), is far more expensive, and subjects patients to much higher levels of ionizing radiation, so it is used infrequently.
MPR is frequently used for examining the spine. Axial images through the spine will only show one vertebral body at a time and cannot reliably show the intervertebral discs. By reformatting the volume, it becomes much easier to visualise the position of one vertebral body in relation to the others.
Modern software allows reconstruction in non-orthogonal (oblique) planes so that the optimal plane can be chosen to display an anatomical structure. This may be particularly useful for visualising the structure of the bronchi as these do not lie orthogonal to the direction of the scan.
For vascular imaging, curved-plane reconstruction can be performed. This allows bends in a vessel to be 'straightened' so that the entire length can be visualised on one image, or a short series of images. Once a vessel has been 'straightened' in this way, quantitative measurements of length and cross sectional area can be made, so that surgery or interventional treatment can be planned.
MIP reconstructions enhance areas of high radiodensity, and so are useful for angiographic studies. mIP reconstructions tend to enhance air spaces so are useful for assessing lung structure.
Surface rendering: A threshold value of radiodensity is chosen by the operator (e.g. a level that corresponds to bone). A threshold level is set, using edge detection image processing algorithms. From this, a 3-dimensional model can be constructed and displayed on screen. Multiple models can be constructed from various different thresholds, allowing different colors to represent each anatomical component such as bone, muscle, and cartilage. However, the interior structure of each element is not visible in this mode of operation.
Volume rendering: Surface rendering is limited in that it will only display surfaces which meet a threshold density, and will only display the surface that is closest to the imaginary viewer. In volume rendering, transparency and colors are used to allow a better representation of the volume to be shown in a single image - e.g. the bones of the pelvis could be displayed as semi-transparent, so that even at an oblique angle, one part of the image does not conceal another.
Some slices of a cranial CT scan are shown below. The bones are whiter than the surrounding area. (Whiter means higher radiodensity.) Note the blood vessels (arrowed) showing brightly due to the injection of an iodine-based constrast agent.
A volume rendering of this volume clearly shows the high density bones.
After using a segmentation tool to remove the bone, the previously concealed vessels can now be demonstrated.
Radiology Medical imaging Neuroimaging
Computertomographie Tomografía axial computarizada OTA Tomographie axiale calculée Tomografi terkomputasi CAT-skanni Tomografia assiale computerizzata טומוגרפיה ממוחשבת komputertomográfia Computertomografie CT Computertomografi Tomografia komputerowa Tomografia computadorizada Компьютерная томография Počítačová tomografia Datortomografi கணிக்கப்பட்ட குறுக்குவெட்டு வரைவி வரிக் கண்ணோட்டம் Bilgisayarlı tomografi X射线断层成像
This article is licensed under the GNU Free Documentation License.
It uses material from the
"Computed tomography".
Home Page • arts • business • computers • games • health • hospitals • home • kids & teens • news • physicians • recreation• reference • regional • science • shopping • society • sports • world