Atomic absorption spectroscopy in analytical chemistry is a technique for determining the concentration of a particular metal element within a sample. Atomic Absorption Spectroscopy can be used to analyse the concentration of over 62 different metals in a solution.
Atomic Absorption Spectroscopy was first developed during the 1950's by a team of Australian chemists, lead by Alan Walsh, working at the CSIRO (Commonwealth Science and Industry Research Organisation) Division of Chemical Physics, in Melbourne Australia. Typically, the technique makes use of a flame to atomize the sample, but other atomizers such as a graphite furnace are also used. Three steps are involved in turning a liquid sample into an atomic gas:
The flame is arranged such that it is laterally long (usually 10cm) and not deep. The height of the flame must also be controlled by controlling the flow of the fuel mixture. A beam of light is focused through this flame at its longest axis (the lateral axis) onto a detector past the flame.
The light that is focused into the flame is produced by a hollow cathode lamp. Inside the lamp is a cylindrical metal cathode containing the metal for excitation, and an anode. When a high voltage is applied across the anode and cathode, the metal atoms in the cathode are excited into producing light with a certain emission spectra. The type of hollow cathode tube depends on the metal being analysed. For analysing the concentration of copper in an ore, a copper cathode tube would be used, and likewise for any other metal being analysed. The electrons of the atoms in the flame can be promoted to higher orbitals for an instant by absorbing a set quantity of energy (a quantum). This amount of energy is specific to a particular electron transition in a particular element. As the quantity of energy put into the flame is known, and the quantity remaining at the other side (at the detector) can be measured, it is possible to calculate how many of these transitions took place, and thus get a signal that is proportional to the concentration of the element being measured.
For a medium temperature flame, acetylene and air is used. A hotter flame can be produced using acetylene and pure oxygen, and an even hotter flame can be attained using nitrous oxide and acetylene, all these mixtures are explosive.
The narrow linewidths of hollow cathode lamps make spectral overlap rare. That is, it is unlikely that an absorption line from one element will overlap with another. Molecular emission is much broader, so it is more likely that some molecular absorption band with overlap with an atomic line. This can result in artificially high absorption and an improperly high calculation for the concentration in the solution. Three methods are typically used to correct for this:
Spectroscopy | Analytical chemistry
Atomspektroskopie | Espectroscopia de absorción atómica | Espectroscopia de absorción atómica | Atomaireabsorptiespectrometrie | Атомно-абсорбционный спектрометр | Atomiabsorptiospektrometri | Atomabsorptionsspektroskopi
This article is licensed under the GNU Free Documentation License.
It uses material from the
"Atomic absorption spectroscopy".
Home Page • arts • business • computers • games • health • hospitals • home • kids & teens • news • physicians • recreation• reference • regional • science • shopping • society • sports • world